Pictures may not be actual SKU

LR26 | EchoPro Radar Liquid Level Sensor Transmitter

Product info
Description
Flowline's LR26 EchoPro LR26 Radar Liquid Level Sensor Transmitter
Configuration Model Codes
Base Model
Series
LR | Radar Sensors
Model
26 | EchoPro LR26 - Liquids - Intrinsically Safe - Flange
Availability Call to confirm
Select model  or  search within  this product range
Models are:  Available to Order
Ships from PA - TX
Typically ships within [ 12 ] days
24/7 Sales & Support 484-544-0506
Headline

EchoPro LR26 Radar Liquid Level Sensor Transmitter

The intrinsically safe 6.3 GHz. pulse radar transmitter provides continuous level measurement up to 114.8’ (35m) with a 4-20 mA analog and HART digital signal output, and is configured via its integral push button display module or HART communicator. The non-contact liquid level sensor is intended for chemical and petroleum storage or process applications in above grade metal or reinforced concrete tanks or below grade tanks of any material. Select this sensor for harsh process conditions with corrosive or low dielectric media, heavy surface foam or agitation, higher pressure, steam, condensation or vapor, and installation in a flange fitting. Due to the wider beam angle, special attention should be given to ensure that the sensor is located with unobstructed measurement space. Typical applications include process tanks.

Features & Benefits
Features:
  • 316L antenna with PTFE inner shield for corrosive media
  • Configuration via push button display or HART communicator
  • 6.3 GHz. pulse radar is unaffected by most foam and agitation
  • 12” (30cm) dead band enables utilization of the entire tank
  • LCD displays level in feet or meters with percent of span bar
  • Optional display mode indicates the echo signal return curve
  • Auto-temperature compensation for accurate measurement
  • Fail-safe diagnostics with selectable signal fail-safe outputs
  • Recognition, storage and rejection of false echo signal returns
  • IP67 stainless enclosure with polycarbonate display window
Additional Notes
Specifications:
  • Range: 12” to 114.8’ (30cm to 35m)
  • Frequency: 6.3 GHz.
  • Accuracy: ± 10mm
  • Dead band: 12” (30cm)
  • Beam angle: 20º
  • Configuration: Push button, HART 7
  • Memory: Non-volatile
  • Display type: LCD, dot matrix
  • Display units: Feet and meters
  • Display bar: Percent of span
  • Display graph: Echo signal curve
  • Supply voltage: 21.6 to 26.4 VDC
  • Max. consumption: 22.5 mA
  • Signal output: 4-20 mA, two-wire, HART 7
  • Signal invert: 4-20 mA, 20-4 mA
  • Signal fail-safe: 3.9 mA, 20.5 mA, 22 mA
  • Process temp.: F: -40º to 266º, C: -40º to 130º
  • Temp. comp.: Automatic
  • Storage temp.: F: -40º to 176º, C: -40º to 80º
  • Pressure: -14.5 to 580 psi (-1 to 40 bar)
  • Enclosure rating: IP67
  • Encl. material: 316L stainless steel w/silicone gasket
  • Encl. window mat.: Polycarbonate
  • Conduit entrance: (1) 1/2” NPT connector
  • Antenna material: 316L w/PTFE inner shield
  • Process mount: – 40: 4” flange (ANSI), – 60: 6” flange (ANSI)
  • Classification: Intrinsically safe
  • Approvals: ATEX IIG Ex ia II C T6…T3 Ga; IEC Exia IIC T6…T3 Ga; CSA Class I, Division 1, Groups A, B, C, D; Class II, Division 1, Groups E, F, G; (T6…T3) Class I Zone 0 AEx ia IIC (T6…T3) Ga Zone 20 AEx ia IIIC (T76ºC to T146ºC) Da Ex ia IIC (T6…T3) Ga Ex ia IIIC (T76ºC to T146ºC) Da
  • Entity parameters: Install per control drawing #301901
  • Certification: FCC
  • Compliance: CE, CRN, RoHS
Dimensional drawings
Call or text: +1 (484) 544-0506

The Supplyline team is here to help!